Operation options
Proline Prowirl F 200 PROFIBUS PA
44
Hauser
8.2.2
Operating philosophy
The individual parts of the operating menu are assigned to certain user roles (operator,
maintenance etc.). Each user role contains typical tasks within the device lifecycle.
Menu/parameter
User role and tasks
Content/meaning
Language
task-oriented
Role "Operator", "Maintenance"
Tasks during operation:
• Configuring the operational
display
• Reading measured values
• Defining the operating language
• Resetting and controlling totalizers
Operation
• Configuring the operational display (e.g. display format, display contrast)
• Resetting and controlling totalizers
Setup
"Maintenance" role
Commissioning:
• Configuration of the
measurement
• Configuration of the inputs and
outputs
Wizards for fast commissioning:
• Set the system units
• Define the medium
• Configure the current input
• Configure the outputs
• Configuring the operational display
• Define the output conditioning
• Set the low flow cut off
Advanced setup
• For more customized configuration of the measurement (adaptation to
special measuring conditions)
• Configuration of totalizers
• Configure the WLAN settings
• Administration (define access code, reset measuring device)
Diagnostics
"Maintenance" role
Fault elimination:
• Diagnostics and elimination of
process and device errors
• Measured value simulation
Contains all parameters for error detection and analyzing process and device
errors:
• Diagnostic list
Contains up to 5 currently pending diagnostic messages.
• Event logbook
Contains event messages that have occurred.
• Device information
Contains information for identifying the device.
• Measured values
Contains all current measured values.
• Analog inputs
Is used to display the analog input.
•
Data logging
submenu with "Extended HistoROM" order option
Storage and visualization of measured values
• Heartbeat
The functionality of the device is checked on demand and the verification
results are documented.
• Simulation
Is used to simulate measured values or output values.
Expert
function-oriented Tasks that require detailed
knowledge of the function of the
device:
• Commissioning measurements
under difficult conditions
• Optimal adaptation of the
measurement to difficult
conditions
• Detailed configuration of the
communication interface
• Error diagnostics in difficult cases
Contains all the parameters of the device and makes it possible to access
these parameters directly using an access code. The structure of this menu is
based on the function blocks of the device:
• System
Contains all higher-order device parameters which do not concern the
measurement or the communication interface.
• Sensor
Configuration of the measurement.
• Output
Configure the pulse/frequency/switch output.
• Communication
Configuration of the digital communication interface.
• Submenus for function blocks (e.g. "Analog Inputs")
Configuration of function blocks.
• Application
Configure the functions that go beyond the actual measurement (e.g.
totalizer).
• Diagnostics
Error detection and analysis of process and device errors and for device
simulation and Heartbeat Technology.