Lynx Genie 2000 Acquisition Window Adjust
Lynx DSA User's Manual - 9240227J
57
MCS Parameters
The MCS controls are accessible only from a MCS datasource. The datasource type
is defined during the MID Editor or MID Wizard session by selecting MCA Number
2 for MCS.
Only MCS setting adjustments are available for the MCS datasource. A PHA
datasource for the device must be opened if adjustments to the Stabilizer, High
Voltage, Gain or Filter are required.
The MCS settings screen for the Lynx contains the following controls.
Figure 29: Adjust Screen’s MCS Settings
Dwell Range
Sets the units for the dwell time value. The unit's selection can be set to
microseconds
,
milliseconds
, or
seconds
.
Dwell Value
Sets the dwell time value (in units specified by the Dwell Range control). The Lynx
supports a dwell value ranging from 1 microsecond to 999 seconds in increments of 1
unit depending on the selected range.
Disc. Mode
The Discriminator Mode control establishes the input signal to the MCS Acquisition
Unit.
•
Selecting
FDisc
enables the MCS’s “fast discriminator mode” where all
incoming events above the Fast Discriminator threshold are counted. This
mode is equivalent to counting the ICR Value reported through the Status
page for the selected dwell time.
•
Selecting
TTL
enables the MCS’s “TTL mode” where all TTL events
received through the rear panel’s MCS input are counted for the selected
dwell time.