Chapter 4 System Setup
48
Lynx DSA User's Manual - 9240227J
Corr. Range
Sets the allowable correction range based on detector type.
NaI
s/- 10% of
full scale,
Ge
s/- 1% of full scale. If the correction exceeds the allowed range
an over-range or under-range error will be asserted by the Lynx.
Ratio Mode
Determines the source of the ratio value to be used for stabilization control.
Manual
uses the value set through the Manual Ratio setting, and
Automatic
allows the Lynx
to calculate the value automatically when acquisition and stabilization are active. For
most typical applications this setting should be set to
Automatic
.
Typical Setup Sequence
For a typical setup sequence for the spectrum stabilizer, please refer to the
Typical
Stabilizer Setup Sequence
on page 113 section.
DSP Gain Parameters
The Gain controls are accessible only from a PHA datasource. The datasource type is
defined during the MID Editor or MID Wizard session by selecting MCA Number 1
for PHA or Number 2 for MCS.
The DSP Gain settings screen for the Lynx contains the following controls.
Figure 23: Adjust Screen’s DSP Gain Settings