Chapter 2 Getting Started
14
Lynx DSA User's Manual - 9240227J
Accessing your Lynx through the UPnP Name
From the Navigator view, select
Maintenance | Network Settings | UPnP
. The
Friendly Name will be displayed. If your network supports UPnP messaging, and
your computer has the UPnP option enabled, then you can access the Lynx anywhere
on the network through its UPnP Friendly Name instead of its IP address.
•
The UPnP Friendly Name can be changed to any custom string up to 64
alphanumeric characters. The entry is restricted only to alphanumeric
characters and a dash (-) character.
•
Factory-reset value for the UPnP Friendly Name is
Lynx
.
Click the
Save
button to save all changes.
Setting the Lynx’s System Time
From the Navigator view, select
Maintenance | MCA Information.
Press
Synchronize
button to push the computer’s timestamp to the Lynx.
The Delta(s) field shows the difference in seconds between the computer’s timestamp
and Lynx’s timestamp.
Viewing the Firmware and Application Versions
From the Navigator view, select
Maintenance | MCA Information
to retrieve the
Firmware and FPGA Versions.
Viewing and Clearing Errors and Faults
From the Navigator view, select
Maintenance | Diagnostics | Status
to retrieve
Errors and Faults. Clear the errors and faults through the
Execute
button.
Please note that if the error of fault condition persists then the indication will
reappear shortly after clearing it.
Setting Graph Preferences
From the Navigator view, select
Preferences | Miscellaneous
or
Axes
or
Colors
to change the graphing properties as desired. Changes take effect
immediately.
Setting PHA Acquisition Parameters
From the Navigator view select the
Acquisition
option and then
Acquisition |
General
to set acquisition preset times,
Acquisition | Coincidence
to set
coincidence parameters, or
Acquisition | External Control
to set external control
parameters.
The Acquisition Mode for Input 1 can be set to PHA, MSS, LIST, and TLIST modes.