Chapter 5 Web-Based Operations
130
Lynx DSA User's Manual - 9240227J
Lock/Unlock
Locks or unlocks the Lynx with the current browser session. Lock provides read and
write access to the Lynx, meaning that the Lynx’s settings and status can be read and
written, data can be acquired, etc. Unlock restricts writing to the Lynx, meaning that
settings and status can only be read and not changed.
When changing state from unlocked to locked the Administrator’s password will be
requested.
Release Notes
The Release Notes feature provides a convenient way to displays the current
version’s release notes for easy reference.
About Application
The About Application feature displays the current version and Copyright date.
Reboot
The Lynx will be rebooted similar to a power cycle. Spectral data for PHA and MCS
will not be affected, however any residual data from Auxiliary Counters, SCA
Counters, List and Time-Stamped List will be lost.
Alternate Web Sites
The Lynx contains a variety of web sites to meet your specific needs. The reasons for
offering a variety of sites to increase the accessibility of the device to modern devices
like tablets and smart phones. Browsers that run on modern devices do not support
plugins. However they support HTML5 that is industry standard for replacing
plugins.
Please refer to
A Brief Overview of the Lynx User Interface
on page 8 for URL’s on
accessing the alternate web sites.
If automatic login is not enabled then you will be presented with the same login
screen for both sites.