Maintenance Options
Lynx DSA User's Manual - 9240227J
129
Restore Settings
To restore acquisition settings press the
Select
button to navigate to the folder
containing the settings XML file where Backup was saved. Select the file and
then click Open to accept the selection. The restore operation typically takes few
seconds.
The Restore operation typically will be disabled if any of the conditions are meet:
Acquire is ON (not Idle state), HV is ON, Stabilizer is ON or on HOLD, or when
the application is in Read-only.
Export Settings
The
Export settings
is used to save all current Lynx settings into a CSV
formatted file by default. The contents include the parameter name, description,
code, and value. Click on the
Select
button and then enter a file name. Click on
OK
to save the settings as a CSV file by default.
Note: This is a Web browser specific setting that will download the file where
the download’s settings location folder is configured to.
Export Spectral Data
The
Export Spectral Data
can be used to save the current spectral data into a
CSV formatted file by default. The contents includes energy calibration, live and
real times, and the channel / energy per counts. Click on the
Select
button and
then enter a file name. Click on
OK
to Click on
OK
to save the settings as a CSV
file by default.
Note: This is a Web browser specific setting that will download the file where
the download’s settings location folder is configured to.
Lynx Reset to Factory Defaults
Select the
Reset
button to reset the Lynx unit to its factory default values. Refer
to Factory Reset Control on page 23.