Chapter 5 Web-Based Operations
116
Lynx DSA User's Manual - 9240227J
Gate Timer (µS)
Sets the Inhibit Gate Delay value by selecting from a list of options ranging from 0 to
700 milliseconds in 100 ms increments; selecting 0 implies no Gate Delay.
Clock Source
The Synchronization Clock Source is enabled only when
External Sync Status
is set
to
Independent
. See
Hardware
|
External Synchronization
option.
If External Synchronization Status control is set to
Independent
, the Clock Source
setting then defines the clock source to be used when external synchronization is
Enabled
. Options are
Internal
or
External
, as follows:
•
Internal
means that the clock source is the internal 1
µ
S clock.
•
External
means that the clock source is defined by the External Sync input at
the rear panel’s SYNC connector.
Enable
Select “On” to enables or “Off” to disable Counter 1 and Counter 2. When enabled,
both counters will count pulses present on the rear panel’s GPI/O 1 and 2 respectively
for precise time intervals set by the Preset Limit setting. Counter data is reported by
the Web Client application through the
Time Series
plot
.
Time Series
Select “On” to enables or “Off” to disables the plot.
External Synchronization Settings
Activates the settings for the External Synchronization parameters.
Figure 61: External Synchronization Hardware Settings