Manual Inhibit Mode
Lynx DSA User's Manual - 9240227J
201
As shown below, the overload recovery time is approximately 45
µ
s, with the Lynx
Rise Time set to 5.6
µ
s, Flat Top set to 0.8
µ
s, Coarse Gain of x304 [find closest
Lynx value] and using a Mirion Model 2101 preamp and
60
Co source. For this
example the Inhibit Time should be set to approximately 45
µ
s.
Figure 107: Digital Oscilloscope Waveform Showing Trapezoid Filter