Chapter 5 Web-Based Operations
98
Lynx DSA User's Manual - 9240227J
•
Single Channel Analyzers
displays the Single Channel Analyzer settings
and associated acquisition control.
High Voltage Power Supply Settings
The High Voltage Power Supply settings include the Range and Polarity selection,
Voltage setting, Inhibit control, On/Off status control, the current high voltage
reading measured by the Lynx, as well as Reset to clear errors and fault conditions.
WARNING
:
High voltage direct current (DC) energy can be present on
the High Voltage Power connectors, which can be harmful.
Please use extreme care.
Figure 53: High Voltage Power Supply Settings
Voltage(V)
The Voltage setting sets the voltage value for the current range. Voltage setting is
limited by the Limit value. Acceptable settings for Voltage range from 0.00 volts up
to the limits’ maximum. When changed, the high voltage output ramps towards its
new destination at a rate of 100 volts/second, updated approximately every 60
milliseconds.
Voltage value can be changed whether the high voltage power supply is
On
or
Off
.
Reading(V)
The Reading is a read-only value reported by the Lynx representing the actual
voltage in volts currently present on the rear panel’s SHV connector. The reading is
updated dynamically every few seconds.