6-6 Thermo
Electron
January
2004
“Write: Could not transmit a packet message. Communications are
probably lost.”
“Write: Could not transmit an ack or nak. Communications are
probably lost.”
“DH Unknown calibration query.”
“DH Unknown calibration query msg type.”
“DH Unknown calibration reset msg. type.”
“Filter Wheel Misaligned!”
CRITICAL ERROR
MESSAGES
Critical Error
messages indicate that the PDA can not perform its
function properly. If a chromatographic run is in progress when a
Critical Error
occurs the data may be corrupted or lost, and the run
will be terminated. Possible
Critical Error
messages include:
“PM: Attach has failed. There may be an existing connection.”
“PM: Queue creation failed; exceeded maximum values.”
“PM: The UV6000LP is running out of internal buffers. Most
probable cause: data rate too high. Injection will be aborted to prevent
message depletion.”
“PM: Could not write to a queue. You should probably exit the
software and/or cycle the power on the UV6000LP to ensure future
data integrity.”
“PM: A queue is too deep, might be a problem. It may be best to exit
the software and/or cycle the power on the UV6000LP to ensure
future data integrity.”
“ PM: Could not transmit; Communications have been lost. You will
have to reconnect .”
“State: Communications have been lost, tossing packets.”
“Tungsten Lamp Failure.”
“Deuterium Lamp Failure.”
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......