3-2 Thermo
Electron
January
2004
Configuring the System
Configure your LC system as you normally would. Because the
UV6000LP connects directly to a serial port of the computer, it must
be manually configured. In the Detector section of the System
Configuration window, select the UV6000LP as the model of the
detector you want to configure. Select the correct COM (serial) port
for the UV6000LP along with both deuterium and tungsten lamps.
Save the configuration.
If you are using the ChromQuest software package, create your LC
Test Mix analytical method as described below. If you are using the
Atlas software package, refer to the Atlas SpectraSYSTEM Control
manual for information on creating instrument methods.
Quick Start with ChromQuest
The following section describes the proper setup of the
chromatographic method for running the Thermo Electron LC Test
Mix with the Microsoft Windows NT/2000/XP-based ChromQuest
software.
Click on the Windows taskbar Start button and move the cursor up
the menu to Programs. Click on the ChromQuest icon located in the
Chromatography program group. Select the appropriate system icon
from the ChromQuest main menu window. The default ChromQuest
windows will open automatically.
INSTRUMENT
SETUP
We begin to create the analytical method by specifying the
operational parameters in the Instrument Setup section. To do this,
select Method-Instrument Setup from the Menu Bar at the top of the
Applications Window. Next, click on the appropriate tab for each
instrument in the system (Figures 3.1 & 3.2) and enter the LC Test
Mix run parameters listed below. It should look like the screens
following this table. When all tabbed pages have been configured,
save the method as LCTestMix.
Pump Program
Solvent(s)
75% methanol, 25% water
Flow rate
1 mL/min
Time
5
min
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......