January 2004
Thermo Electron
5-29
Turn Off Button
Click on this button to turn off the Event Output state.
Deuterium Lamp
Lamp On Checkbox
Select this checkbox to turn on the deuterium lamp and deselect this
checkbox to turn off the deuterium lamp. The deuterium lamp takes
approximately 30 seconds to ignite and approximately 60 minutes to
warm-up.
Deuterium Lamp Lifetime Read-back
This read-back displays the passage of time in hours that the
deuterium lamp has been turned on since the last time the Reset
button was clicked. (The Reset button allows you to zero the read-
back. Click on the Reset button after you install a new deuterium
lamp.) The maximum lifetime of the deuterium lamp is typically
2000 hours. In general, as the hours of usage increases, the baseline
noise increases.
Reset Button
After you replace the deuterium lamp, click on this button to reset the
time clock to zero.
Tungsten Lamp
Lamp On
Select this checkbox to turn on the Tungsten lamp and deselect this
checkbox to turn off the Tungsten lamp.
Tungsten Lamp Lifetime Read-back
This read-back displays the passage of time in hours that the tungsten
lamp has been turned on since the last time the Reset button was
clicked. (The Reset button allows you to zero the read-back. Click on
the Reset button after you install a new tungsten lamp.)
Reset Button
After you replace the tungsten lamp, click on this button to reset the
time clock to zero.
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......