6-4 Thermo
Electron
January
2004
Troubleshooting Table (continued)
Symptom Cause Remedy
8. Flat baseline, portion of
spectrum missing.
a. Saturation of photodiode
array.
a. Adjust attenuators.
10. Status LED States
a. Red - Power On, performing
post.
a. No action required.
b. Flashing Red - Failed POST
or fatal error afterwards.
b. Power-on Self Test.
c. Flashing Green - Passed
Post.
c. Needs firmware download.
d. Green - OK, software
loaded, ready to connect to
PC.
a. No action required.
Error Messages
For information on error messages that occur as part of the
UV6000LP diagnostics program, refer to the Log Entries section
below.
Log Entries
The diagnostics Error Log page chronologically records messages
relating to detector/system problems. There are three categories of
messages;
Warning
messages,
Critical Error
messages and
Crash
Error
messages.
WARNING
MESSAGES
Warning
messages indicate a problem that should not affect the
chromatographic run. Possible
Warning
messages include;
“PM: A message is improperly addressed.”
“PM: Free buffers back to acceptable level. You may have to lower
your data rate to continue without further errors.”
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......