January 2004
Thermo Electron
5-27
Spectrum Stop Text
Box
The Spectrum Stop text box allows you to specify the end of the scan
range. The allowable values are 191 nm to 800 nm in the absorbance
mode and diode number 3 to diode number 511 in the intensity mode.
Channels A/B/C
The channel text boxes allow you to specify a wavelength value from
190 nm to 800 nm or a diode number from 2 to 511.
Start Button
Clicking on the Start button starts the collection and display of data
from the detector.
Stop Button
Clicking on the Stop button ends the collection of data.
Zero Button
Clicking on the Zero button clears the displays and empties the data
buffers of the detector.
THE MAINTENANCE
PAGE
The Maintenance page allows you to turn the lamps on and off,
perform a wavelength calibration and an array calibration, turn the
Ready Outputs on or off, and turn the Event Outputs on or off. See
Figure 5.21.