Troubleshooting Table (continued)
Symptom Cause Remedy
2. Random noisy baseline.
a. Contaminated flowcell.
a. Flush flowcell with cleaning solvents
as described in Chapter 3. Check for
leaks.
b. Leak in sample inlet line.
b. Check all fittings from column outlet
to flowcell inlet for leaks.
c. Bubble trapped in flowcell.
c. Increase flow rate until bubble is
removed. Supply back-pressure
device to flowcell (check back-
pressure rating to avoid rupturing
flowcell).
d. Leaking flowcell.
d. Replace flowcell.
e. Insufficient lamp warm-up.
e. Allow a 30 minute warm-up for
normal operation; 11/2 hours for
maximum sensitivity.
f. Lamp aging or defective.
f. Replace lamp.
g. Ground loop problem
between integrator and
detector.
g. Check for proper cable connections
for detector output; do not ground at
both ends of cable.
h. Flowcell or lamps dirty.
h. Clean dirty component as described
in Chapter 3.
i. Integrator input voltage does
not match detector output
voltage.
i. Connect integrator to appropriate
output connectors on detector. Check
attenuation setting on integrator.
3. Excessive baseline drift,
see
Baseline Problems
a. Flowcell contaminated.
a. Flush flowcell with cleaning solvents
as described in Chapter 3. Check for
leaks.
b. Mobile phase contamination. b. Replace with fresh mobile phase
made with high-purity solvents.
c. Material bleeding from
column.
c. Clean or replace column.
d. Leaks in system, or flowcell. d. Check all fittings for leaks. Replace
flowcell.
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......