1-2 Thermo
Electron
January
2004
Chapter 3, UV6000LP Quick Start -
Contains example procedures for
running a standard on the UV6000LP using ChromQuest Ver. 4.1.
Chapter 4, Routine Maintenance
- Recommends and includes
procedures for routine maintenance to keep your detector in peak
operating condition.
Chapter 5, Diagnostics
- Describes the user interface and use of the
UV6000LP Diagnostics program supplied with ChromQuest Ver. 4.1
and the Instrument Manager supplied with Atlas Ver. 2003, R2.
Although this is part of the software, it is included here rather than in
the software manual since it is specific to this detector. Several
diagnostic procedures are recommended as routine maintenance in
Chapter 4.
Chapter 6, Troubleshooting
- Contains a table of symptoms, possible
causes, and remedies for some common problems you may observe
when using your detector.
Chapter 7, Accessories and Replacement Parts
- Lists accessories and
user spare parts for the detector and includes Thermo Electron part
numbers.
Chapter
8,
Field Service Procedures
- provides detailed instructions
for factory trained service personnel to perform field repairs.
UV6000LP Functional Description
This brief functional description of the UV6000LP is included to
assist you with troubleshooting potential problems and performing
routine maintenance on your detector. For more detailed information,
contact your Thermo Electron representative.
The UV6000LP Detector is packaged as a bench-top unit for
inclusion in a liquid chromatography system. It is designed to be
remotely operated over a serial communications link from a PC using
either Microsoft Windows NT/2000 based ChromQuest or Atlas
software. The detector consists of a dual-light source, an optical
bench, a photodiode array, a low voltage power supply, and several
printed circuit board assemblies (PCBAs).
Figure 1.1 shows the optical system used in the detector. The dual-
light source includes a deuterium lamp for detection in the ultraviolet
wavelength range (190 nm to 350 nm) and a tungsten - halogen lamp
for detection in the visible wavelength range (350 nm to 800 nm).
There is some overlap between the two lamps in the 300 nm to
500 nm range.
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...