4-16 Thermo
Electron
January
2004
6.
To remove a lamp assembly, first disconnect its connector from
the lamp power supply board by holding the connector and
gently pulling the connector free from the contact pins on the
PC board.
7.
Next loosen and remove the single Phillips head screw securing
the tungsten - halogen (W) lamp mounting block in place. In
the case of the deuterium (D2) lamp, loosen (but do not
remove) the two Phillips head screws, lift and gently twist the
mounting flange to free the lamp.
8.
Lift out the wiring between the connector and lamp socket
from the compartment.
To install a lamp, perform the above removal procedure in reverse
order.
NOTE: The surfaces of both lamps must be kept free of fingerprints and
smudges. If either lamp needs cleaning, use a lint-free lens paper moistened
with methanol or isopropanol before replacing the lamp cover.
NOTE: When a lamp is replaced don’t forget to reset the lamp hours.
(Refer to the procedure for ChromQuest users on page 5-3 or the procedure
for Atlas users on page 5-29.)
Adjusting Attenuators
During the lifetime of the UV6000LP it may be necessary to
manually adjust the attenuators in order to increase the amount of
light falling upon the array or to decrease the light levels to prevent
saturation of the array. For optimal performance the light intensity
values should fall within the following ranges in the region where
analyses will be performed:
•
Units with S/N
≤
107999: 300,000 to 650,000 counts
•
Units with S/N
≥
109000: 300,000 to 800,000 counts
Light intensities should be within the above ranges to minimize
baseline noise and prevent saturation of the diode array. (If the array
is saturated the response from the UV6000LP will be a flat baseline.)
Attenuators may require adjustment when either of the lamps are
replaced, or when a flowcell is replaced. When these items are
replaced, check the light intensities by following the procedure on
page 4-6 (Operation Verification) and adjust the attenuators to
provide light intensities in the specified operating ranges.
As lamps age their light intensities will decrease slightly. If a
significant drop in light intensity occurs, the attenuators can be
opened up to compensate and increase the light intensity.
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......