January 2004
Thermo Electron
2-5
3.
If necessary, remove the voltage selection PC board (it slides
out) and install it with the desired voltage readable. This
voltage must agree with the incoming line voltage.
4.
Verify that the fuse holder contains the appropriate size fuses.
5.
Slide the fuse holder back into the power entry module until it
snaps in.
NOTE: Do not connect the power cord and turn on the detector yet. Wait
until the Power On section of this chapter.
Making Initial Rear Panel Connections
Use the cables supplied with your detector to make the connections
described in this section (see Figure 2.2). For each connection, insert
the cable’s bare wire into the appropriate detector terminal. Hold the
wire in place while you tighten the small terminal set-screw firmly
onto the wire.
PD
-Z
003E
1
2
3 4 5 6
ch 1
ch 1
ch 2
ch 2
1 - 1000 mv - 4
2 - 100 mv - 5
3 - 10 mv - 6
OUTPUT
INPUT
START
ZERO
SPARE
COM
COM
COM
Serial No.
UV6000LP
Model:
5V
EVENT
READY
RUN
SPARE
COM
COM
COM
RS232
+
-
+
-
Figure 2.2 Detector rear-panel connections