January 2004
Thermo Electron
3-7
RUN SAMPLE
When the baseline is stable, click the Stop button on the Command
Toolbar to discontinue monitoring the baseline. To inject the LC Test
Mix sample, click on the Run Sample button on the Command
Toolbar. The Single Run Acquisition dialog box will open requesting
information regarding this particular injection (Figure 3.5).
Enter the following information in the Single Run Acquisition
dialog box:
Data file:
LCTestMix
Vial:
A01
Injection volume:
10 µl
Select the Print Method Report check box. Other fields can be left
blank or at their default settings.
Figure 3.5 Starting a Single Analysis
When the sample is injected, the chromatogram(s) will be displayed
in real-time. If the chromatograms for the UV6000LP are not
displayed, verify that there is a run/start connection between the
autosampler and the UV6000LP. (Because the UV6000LP plugs
directly into the PC, it does not receive a start command from the
SN4000 controller and must be “hard-wired” to the AS3000
autosampler.) At the end of the run the reports that were selected in