4-8 Thermo
Electron
January
2004
Figure 4.2 Intensity Scan with D
2
Lamp
14.
Figure Click on the Status-Off and Data-Stop buttons. Go to
the Lamps page and turn the tungsten - halogen (W) lamp on.
15.
Return to the Control page and changes the individual channels
to monitor pixels 94, 177 and 260.
16.
Go to the Display page and once again click on the Status-On
and Data-Start buttons. See Figure 4.3. Use the Print Screen
key to print out a copy of the scan. This scan should also be
kept for future comparisons to see if there is any intensity
degradation. Date the printout and add it to your maintenance
records.