![Thermo Finnigan SpectraSYSTEM UV6000LP Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/thermo/finnigan-spectrasystem-uv6000lp/finnigan-spectrasystem-uv6000lp_reference-manual_1104224052.webp)
3-8 Thermo
Electron
January
2004
the Method Custom Report will print automatically. The system will
then go to an idle state.
EXPECTED
RESULTS
The table 3.1 below shows the typical retention times and peak areas
for the compounds in the LC Test Mix. Typical peak areas for a
SpectraSYSTEM UV1000 detector are also included to show the
increased sensitivity of the UV6000LP.
Table 3.1 Expected Results for LC Test Mix
Injection Amount
(ng)
Compound Ret.
Time
(min)
UV1000
Peak Area
UV6000
Peak Area
60
Tartrazine
0.61
150908
723005
20 Acetophenone
1.48
23336
119167
1000 Toluene
2.07
103254
526428
5 Anthracene
3.60
139472
688661
VIEW REPORT
In addition to viewing the analysis on-screen in the discrete
wavelength windows and the spectral analysis window, the complete
LC Test Mix analysis is printed in the Area % Custom Report format.
The report can be viewed by selecting Custom Report from the
Method pull-down menu. The Method Custom Report window opens
with your LC Test Mix analysis data.
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......