January 2004
Thermo Electron
4-11
Figure 4.5 UV6000 Instrument Status – Maintenance page
6.
Click on the Display tab to open the Display page. See
Figure 4.6. (This tab is labeled
Operation
in Atlas version
2003 R2.)
7.
Select the Diode Number Intensity option button.
6.
Set the individual channels to monitor diodes 35, 52, and 77.
Then, click on the Start button.
7.
Return to the Maintenance page and turn the tungsten - halogen
(W) lamp on.
8.
Return to the Display page and set the individual channels to
monitor diodes 94,177, and 260. Then, click on the Start
button.