January 2004
Thermo Electron
8-5
CPU PCBA
Use the following procedure to remove the CPU PCBA:
1.
Disconnect the eight connectors across the top of the PCBA,
(Figure 8.3) For the three ribbon-cable connectors, open and
spread the retaining clips prior to removing.
2.
Using the flat of a small screwdriver blade gently push the two
black snap latches free from their sockets at the top corners of
the PCBA.
3.
Lift the PCBA out of the chassis.
PD-Z1
10E
JP16 to
Array Acq.
PCBA
JP7 to
Front Panel LED
JP10 to
Filter Wheel Sensor
JP1 to
Main Power Supply PCBA
JP5 to
Lamp Power Supply PCBA
P7 to
Chassis Ground
JP7 to
Transition PCBA
Figure 8.3 CPU PCBA Connector Layout