January 2004
Thermo Electron
3-5
and the bandwidth of the scan data can be set appropriately to
maximize spectral resolution. This is extremely important when
comparing detailed spectra for compounds at very low concentrations.
NOTE: ChromQuest can automatically display three signals simultaneously.
The UV6000LP configured as the only detector in an LC system can monitor
three signals simultaneously, which are selected in the discrete channels
section of the UV6000LP acquisition method page. If another detector is
configured in the system, the UV6000LP will automatically be set up to
monitor two channels. Other signals can be added after the run starts. If
scan data is collected, individual chromatograms at other wavelengths can
be viewed after the run.
CUSTOM REPORT
The ChromQuest Custom Reporting feature allows the analytical
chemist to consolidate test results into a single report with a
maximum amount of flexibility. To simplify the process,
Thermo Electron has included four of the most common report
formats as report templates. Additional reporting capabilities are
described in the
ChromQuest Software User’s Guide
.
To activate automatic reporting, select Custom Report from the
Method pull-down menu. The Method Custom Report Window will
open. Next, click on File-Report Template-Open and select the area
% template from the ChromQuest folder by double-clicking on the
file Area%.Srp (Figure 3.3).
Figure 3.3 Custom Report Template
The template will now appear in the Method Custom Report Window.
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......