January 2004
Thermo Electron
2-13
Calibrating the UV6000LP
The UV6000LP was calibrated at the factory; however, it should be
recalibrated after it is installed and before it is used for analysis. The
calibration is dependent on the exact physical alignment of
components on the optical bench within the unit. Bumps and jars,
which might occur in shipping and during the installation process, can
affect this alignment.
Easy calibration procedures are included in the UV6000LP
Diagnostics section. Refer to Chapter 5 for information on performing
the wavelength and dark current calibration procedures.
Operation Verification
Once the detector has been successfully installed, refer to
Operation
Verification
in Chapter 4 to verify that the detector is working
properly and to establish initial emission intensity spectra of
the lamps.