January 2004
Thermo Electron
2-3
Making Initial Preparations
Place the detector on a bench top as close as possible to the
chromatographic column outlet (This minimizes the length of tubing
necessary for connection to the flowcell inlet). Be sure to place the
detector in a draft-free location away from an opened window, air
conditioner vents or other circulating air source. A stable room
temperature is necessary for applications requiring maximum
detection sensitivity. Allow at least five inches (13 cm) of clear space
between the detector’s rear panel and any wall or obstruction. This
provides access to the rear-panel connectors and a free flow of
cooling air.
You will need the following tools for installation:
•
a narrow-tip screwdriver (2 mm wide)
•
a #2 Phillips screwdriver
•
a ¼
″
open end wrench