January 2004
Thermo Electron
6-3
Troubleshooting Table, continued
Symptom Cause Remedy
3. Excessive baseline drift,
see
Baseline Problems
(cont.)
e. Tiny bubble trapped in
flowcell.
e. Increase flow rate until bubble is
removed. Connect back-pressure
device to flowcell outlet (check back-
pressure rating to avoid rupturing
flowcell).
f. Large temperature
fluctuations.
f. Remove system from drafts.
Thermostatically control column
temperature.
4. No peaks, or peaks much
smaller than expected.
a. Incorrect wavelength
setting.
a. Check wavelength setting. Make sure
the correct file is selected.
b. Lamp not on, or defective.
b. Make sure lamp is lit. Run detector’s
diagnostic tests to check lamp.
Replace lamp if necessary.
c. Integrator input voltage does
not match detector output
voltage.
c. Connect integrator to appropriate
output connectors on detector. Check
attenuation setting on integrator.
d. Insufficient sample reaching
the detector.
d. Check entire chromatographic system
for leaks. Verify sample injection
volume.
5. Broad, tailing peaks.
a. Rise time is too large (too
slow).
a. Lower the rise time selection.
b. Poor connection at flowcell
inlet.
b. Check end of inlet tubing for a clean,
flat surface free of obstructions.
6. Detector will not
power up.
a. Tripped circuit breaker at
AC wall outlet.
a. Resolve problem, reset circuit breaker.
b. Blown detector fuse.
b. Resolve problem, replace fuse.
c. Incorrect voltage selected.
c. Reset detector for correct incoming
line-voltage (see Chapter 1).
d. Power cord not connected. d.
Connect
power
cord.
7. Detector does not go into
run upon injection.
a. Detector not receiving
START signal.
a. Check connection to Run contacts on
back panel.
Содержание Finnigan SpectraSYSTEM UV6000LP
Страница 1: ...Finnigan SpectraSYSTEM UV6000LP Detector Reference Manual A0099 595 Revision F ...
Страница 8: ......
Страница 12: ......
Страница 16: ......
Страница 44: ......
Страница 76: ......
Страница 91: ...January 2004 Thermo Electron 5 15 Figure 5 8 Holmium oxide absorbance spectrum ...
Страница 118: ......
Страница 129: ...January 2004 Thermo Electron 8 11 PD Z115E Figure 8 6 Power button and bracket removal ...
Страница 134: ......