4-12 Thermo
Electron
January
2004
9.
Return to the Maintenance page and turn the deuterium (D
2
)
lamp off. This will allow you to view the intensity spectrum of
the Tungsten (W) lamp alone.
10.
Return to the Display page and set the individual channels to
monitor diodes 219, 302, and 427.
11.
Return to the Maintenance page and turn the deuterium (D
2
)
lamp on again. Exit the Instrument Manager application and
allow sufficient warm-up time before acquiring data.
Figure 4.6 UV6000 Instrument Status – Display page (Atlas 2003, R1)