6.4.6
Firewall and Network ports
The installation contains several modules (see chapter 6.5). Communication between
modules is performed via network components (Sockets, Ports). Ports that are
utilized are listed in the following table. The Firewall must open the ports and let the
program module communication pass through the Firewall.
Transwin32.exe
Port 127.0.0.1:23945
ServiceTool.exe
Port 127.0.0.1:21947
MorpheusHost.exe
Port 127.0.0.1:21946
MLV_Control_PC.exe
Port 127.0.0.1:10061
VeriCube_Control_PC.exe
Port 127.0.0.1:10051
Morpheus Broadcast
PC-IP/device-IP: 10049
readb.exe (Article database) n/a
If any of the program modules or ports are blocked then the application will not work
or show odd behavior.
6.4.7
Using a gateway
The REA VeriCube is using the camera protocol GigE Vision. This is primarily specified
for a direct communication between two devices (PC and Camera). Operation over a
Gateway is not recognized and does not work.
6.4.8
Fixed IP Addresses
In some network environments fixed IP Addresses are demanded or required. This is
usually solved by a manual configuration of the network settings of the PC and the
VeriCube. A manual network configuration restricts the flexibility of use of the
verifier. Therefore the use of a DHCP server is always recommended. The DHCP
server manages devices which are identified by their MAC addreses. For each device
requiring a fixed IP address the server is configured to provide such devices always
and automatically the same (fixed) IP address.