© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 82 of 119
Also the two transformation options shall be enabled for DPM codes only. This setting
is not according to ISO/IEC 15415.
Coordinate Transformation:
Activating this function is useful for 2D codes applied to cylindrical parts. A distortion
of the codes caused by the radius of the part is compensated by coordinate
transformation. It is planned to use this function for checking in production, as
inaccuracies in the alignment of the code may occur. The approval of printing
layout/laser marking should be based on a measurement without activating the
coordinate transformation.
Reflectance Transformation:
If enabled, bright spots on very glossy metal material will be ignored for verification.
12.2
Symbology selection
The symbology selection determines which of the symbologies and with which
respective properties they are verified. The selection and the properties determine
the respective use.