9.1.2
Reflectance adjustment
Please use calibration card (90031702) for this step.
After level correction is finished, the menu item “reflectance adjustment” is
automatically selected. However, it can be switched in between the different types of
adjustment by selecting the appropriate icon.
The device must be positioned with the help of the live video representation in such a
way that the red cross hair is in the middle of the picture of the four reflectance
fields, which are in the middle of the "Reflectance calibration field" field. The software
automatically detects in which quadrant the bright field lies so that the direction is
not relevant.
Picture: Cross hair in the gray value field
Then the Reflectance adjustment is started with the <OK> button.