© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 67 of 119
11.2
Measuring process using the Windows software.
A prerequisite for performing the measurement is the observation of the handling
instructions (see section 7) as well as the selection of the desired position of the
device (see section 11.1)
The live video is turned on for the positioning. For this purpose, the F1 button on the
device can be used or the symbol button for the live video on the PC screen.
Start the live video
Stop the live video
The measurement area can be set in various ways. Most simply, only the red cross
hair is displayed. The code is moved in the middle of the cross hair and then the
measurement is started. To start the measurement, either the "Scan" button on the
device or the "Evaluation" button in the Windows software can be used.
Cross hair in the barcode
Cross hair in the 2D code
The evaluation is started with the "VERIFY" button.
Pushing the evaluation button stops the live video and starts taking the measuring
image in the full camera resolution. Afterwards, the evaluation is initiated. In a
normal case, the symbology and the code type (1D or 2D) are recognized
automatically.
If the automatic recognition fails, the measurement area must be set manually.