REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 41 of 119
If the function “Only evaluate images with valid adjustment and calibration” is not
activated, the user only receives a warning when the calibration has expired. This
information is then also printed on the test report.
Note: this setting is saved individually in each profile
The instructions for performing a correct measurement must be observed particularly
closely during calibration (see section 7.2).
The adjustment and calibration data of the REA VeriCube are, in contrast to
the other REA Verifiers, saved in the PC – not on the device.
Therefore, software installations on other PCs must always be adjusted
again or the adjustment data must be imported from the first PC
(see section
12.6) if the device is swapped to another PC.
The REA calibration cards are used for adjustment and calibration.
The following instructions for handling the calibration cards must be observed.
The calibration cards must not be dirty or scratched and their surface should
not be touched.
The card for adjustment of the white level correction is protected in the
delivered condition with a blue plastic foil. It must be removed prior to
adjustment.