6.6
Mounting and exchanging the camera lens module
Four different camera lens optic modules can be obtained for the testing of very
small or very large codes. The optic modules are interchangeable and can be
exchanged by the operator without the need for any tools. The different variants with
focal length, minimum code size and field of view size are described in section 2.2.
When changing the camera lens module, the device must be turned off. The knurled-
head screw is unscrewed and the module can be removed.
Picture of an camera lens module
The black dust-protection cap on the lens must be removed prior to mounting. For
each optic module type, one calibration dataset is saved. When a given camera lens
module is used for the first time, the calibration must be performed prior to any
measurement.
ATTENTION: The lens of the optic module is optically measured and is
preset. The alignment shall not be changed and the lens shall not be
mounted on other electronics.