4-24 VS 9/VS 8 Vital Signs Monitor Service Manual
4.9.9
Software Version
Select
Main Menu
quick key
→
from the
System
column select
Version
to show software
version information.
Maintenance Test Report
(See the above sections for detailed test procedures and contents.)
Customer name
Customer address
Servicing person
Servicing company
Equipment under test
(EUT)
Model of EUT
SN of EUT
Hardware version
Software Version
Test equipment
Model/No.
Effective date of calibration
Test Item
Test Record
Test results
(Pass/Fail)
Visual Inspection
The enclosure, display screen, buttons, rotary encoder,
power cord, wall-mount bracket and accessories have
no obvious signs of damage.
The external connecting cables are not frayed and the
connector pins are not loose and bent.
The external connectors are not loose or their pins are
not bent.
The safety labels and data plate are clearly legible.
Power-on test
The power-on test is passed. The power indicator and
alarm system work correctly, and the monitor start up
properly.
Performance Test
SpO
2
test
Measure SpO
2
on a healthy person's finger and a Pleth
wave and PR value are displayed. The displayed SpO
2
value is within 95%-100%.
NIBP Test
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...