6-4 VS 9/VS 8 Vital Signs Monitor Service Manual
6.3
(A) Disassembly of VS9 Main Unit
NOTE
Make sure that the AC power is disconnected and the monitor is off.
Before disassembly, make sure that the point for placement is smooth and
free of unrelated things; otherwise, the screen or the knob may be
scratched or damaged.
All the operations must be performed by professionals. Be sure to wear
insulating gloves during the repair.
When optional functions are indicated, the related operations may be
involved if this function is selected for the machine; otherwise, the related
operations are not involved.
6.3.1
(A-1) Body Temperature Module
NOTE
Different body temperature modules can be removed in the same way. The
quick body temperature module is used as an example here.
1. Use a Phillips screwdriver to unscrew the 2 ST3.3*8 tapping screws at the position
shown in the figure on the rear cover.