1-2 VS 9/VS 8 Vital Signs Monitor Service Manual
1.2.2
WARNING
WARNING
Only personnel authorized by Mindray can disassemble the monitor.
To avoid electric shock, disconnect the AC power supply and remove the
battery before disassembling the monitor.
Perform the patient leakage current test on a reassembled parameter
module before using it for monitoring.
Connect the monitor only to a properly installed power outlet with
protective earth contacts. If a protective earth conductor is not provided,
operate it on battery power, if possible.
Dispose of package materials as per the local laws or applicable waste
control regulations of the hospital, and keep them out of children's reach.
1.2.3
CAUTION
CAUTION
When performing performance tests on the monitor, make sure the
monitor is not exposed to electromagnetic radiation. Mobile phones, X-ray
equipment or MRI devices are possible sources of interference as they may
emit higher levels of electromagnetic radiation.
Before connecting the monitor to the power line, check that the voltage
and frequency ratings of the power line are the same as those indicated on
the monitor's label or in this manual.
Protect the monitor from damage caused by drop, impact, strong vibration
or other mechanical force.
Use only parts and accessories specified in this manual.
1.2.4
NOTE
NOTE
Refer to Operator's Manual for detailed operation and other information.
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...