A-14 VS 9/VS 8 Vital Signs Monitor Service Manual
Symptom
Possible Cause
Recommended Measures
restarted, return this monitor to Mindray
for repair.
Static IP addresses
conflict.
Check if the monitor displays a message
about an IP address conflict.
Check whether the same IP address is
assigned to multiple sets of equipment.
Multiple monitors
are disconnected
intermittently.
APs in some areas are
damaged.
Ensure that the APs have been turned on
and operate properly.
There is strong
interference in some
areas.
Use the network survey tool to check if
the interference is strong. Remove
obvious interference sources or adjust
the WLAN deployment to meet the
requirements of Mindray.
Signal coverage is
insufficient in some
areas.
Use the network survey tool to check the
signal coverage. If the signal coverage is
insufficient in a certain area, adjust the
location of the AP or add APs.
All the monitors
are disconnected
intermittently.
The wired network is
not configured
properly.
Use the wired monitor to check the
wired network configuration. Ensure that
the WLAN bandwidth configured on the
switch is sufficient, with a 50% surplus.
There is radio
interference.
Use the network survey tool to check if
there is any radio interference. Remove
obvious interference sources or adjust
the WLAN deployment to meet the
requirements of Mindray.
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...