VS 9/VS 8 Vital Signs Monitor Service Manual
8-9
Tap
on the touchscreen or press the
arrow key on the
keyboard to switch the upgrade program upward.
Tap
on the touchscreen or press the
arrow key on
the keyboard to switch the upgrade program leftward.
Tap
on the touchscreen or press the
arrow key on the
keyboard to switch the upgrade program rightward.
Tap
on the touchscreen or press the
Enter
key on the
keyboard to confirm the selected upgrade program.
8.2.3.4
Completing Upgrade
When the screen below is displayed, the current upgrade is completed. The upgrade
takes effect after the monitor is restarted.
CAUTION
Disconnect the monitor from the patient and make sure that important
data are saved before upgrade.
Do not shut down or power off the equipment when upgrading the BIOS
program and FPGA program. Otherwise, the equipment may break down.
Program upgrade can be performed only by qualified service personnel.
Indicates unsafe service practice that, if not avoided, could result in slight
injury, product fault, or property loss.
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...