6-38 VS 9/VS 8 Vital Signs Monitor Service Manual
6.4.5.1
(F-1) Removing the NIBP Valve
Unplug the six-way air tube 1 from the valve assembly, and take the valve out of the
buckle structure 2 to remove the valve assembly (note the installation position of the fast
and slow valve during installation).
6.4.5.2
(F-2) Removing the NIBP Pump
Cut off the two cable ties that fix the pump at position 1 and unplug the air tube 2 to
replace the inflation pump and NIBP pipeline components separately. When installing the
pump cable, wind the cable around the column marked by the red circle in the figure.