4-12 VS 9/VS 8 Vital Signs Monitor Service Manual
6. Open the valve to start flowing CO
2
and make sure that there is sufficient flow to
vent to atmosphere.
7. In the text box on the upper right corner of the
Calibrate
menu, enter the vented
CO
2
concentration. The measured CO
2
concentration is displayed in the
Calibrate
menu.
8. After the measured CO
2
concentration becomes stable, select
Calibrate
to calibrate
the CO
2
module.
If the calibration succeeds, the message
Calibration Completed!
is displayed in the
Calibrate
menu. If the calibration fails, the message "
Calibration Failed!
" is displayed. In
this case, perform calibration again.
Troubleshooting for Calibration Failed
Error Code or Failure
Description
Cause
Handling Method
Calibration fails
Module measurement deviation
exceeds the allowable range.
Replace the CO
2
module.
4.5.5
Nurse call relay performance test
This test mainly tests the nurse call function.
4.5.5.1
Test Preparations
Test tools:
Oscilloscope
4.5.5.2
Procedure
1. Connect the nurse call cable to the multifunction output connector of the monitor.
2. Select
Main Menu
->
Maintenance
-> enter the user maintenance password ->
Alarm
->
Nurse Call
.
Monitor
Cylinder
Relief valve
T-shape connector
Tubing
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...