6-20 VS 9/VS 8 Vital Signs Monitor Service Manual
6.3.8.1
(G-1) Removing the AC-DC Power Board
Unscrew the 4 M3*6 pan head screws with washers at the position shown in the figure on
the rear cover to replace the power.
6.3.8.2
(G-2) Removing the AC Socket
Remove the black socket silicone washer as shown in the figure. Unscrew the 2 M3*6
countersunk head screws, and finally unscrew the 1 M4*8 combination screw to replace
the VS9 AC socket cable.
6.3.8.3
(G-3) Removing the Connector Board
Unscrew the 2 M3*6 pan head screws with washers at the position shown in the figure to
replace the VS9 connector board.