4-8 VS 9/VS 8 Vital Signs Monitor Service Manual
button on the right side of the menu to start the test. When the test succeeds, the
NIBP menu will display the message of
Test Successful
. If the pressure exceeds
160-165 mmHg, the message
Test Failed
will be displayed.
4.5.3
Temperature Test
4.5.3.4
SmarTemp Module Test
Required Tool:
Thermostatic oil tank (HART 7102 recommended) or Water Bath and Precision
Thermometer for reference
Test procedure:
1. Set the temperature of the oil tank or water bath to 37°C and conduct the test after
the temperature stabilizes.
2. Set temperature unit to °C.
3.
Select P/M in the
temperature parameter area
and set the measurement mode to
M
(which stands for
Monitor)
.
4. In Monitor Mode, remove the Temperature probe from the probe sheath, insert a
probe cover, and place the probe into the oil tank or water bath.
5. Wait till the Temp value displayed on the monitor stabilizes. Verify that the displayed
value is ±0.1°C of the oil bath thermostat setting or reference thermometer.
Contact our Technical Support Department if the Temperature test fails.
NOTE
Due to the different application environment and the test object in vivo
and vitro conditions, there are deviations in the measurement result. The
maximum deviation of 2.5°C may exist in predictive Temperature
measurement by liquid bath techniques.
4.5.3.5
Exergen Thermometer Verification Test
Calibration can be verified in the lab or clinical units quite easily using a device known as
a portable blackbody. A portable blackbody is a reference heat generator (Figure 1),
which is a self-contained device providing a stable reference target temperature in the
clinical temperature range.
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...