VS 9/VS 8 Vital Signs Monitor Service Manual
6-23
6.3.10.4
(E-3-1) Removing the Display Cable
Remove the corresponding socket and back tape of the display cable indicated by the
arrow, and remove the tape fixing the touchscreen FPC to replace the display cable.
6.3.10.5
(E-3-2) Removing the Display Assembly
Unscrew the 10 M3*6 pan head screws with washers as shown in the figure, and remove
the touchscreen FPC fixing tape and adhesive to remove the display assembly.