VS 9/VS 8 Vital Signs Monitor Service Manual
4-5
Manufacturer
SpO
2
Pulse Rate
Nellcor
96% ± 2% (Adult,
pediatric) 96% ± 3%
(Neonate)
80 ± 3 bpm
Masimo
96% ± 2% (Adult,
pediatric) 96% ± 3%
(Neonate)
80 ± 3 bpm
4.5.2
NIBP Test
4.5.2.1
NIBP Leakage Test
Test the NIBP module and accessories for leaks.
Test Preparations
Test tools:
NIBP cuff for adult patient
Air tubing
Cylinder
Procedure
Follow this procedure to perform the test:
1. Set
Patient Category
in
Patient Information
to
Adult
on the monitor.
2. Connect the NIBP cuff to the NIBP connector on the monitor.
3. Wrap the cuff around the cylinder as shown below.
4. Select
Main Menu
->
Maintenance
-> enter password ->
Module
->
NIBP
->
NIBP
Leakage Test
. The message
NIBP Leakage Test
is displayed in the NIBP parameter
area.
After dozens of seconds, the cuff automatically deflates, which indicates that the NIBP
leakage test is completed. If no message is displayed in the NIBP parameter area, it
indicates that the system has no leakage. If the message "
NIBP Pneumatic Leak
" is
displayed, it indicates that the system may have a leakage. Check the tubing and
Monitor
NIBP cuff connector
Air tubing
Cylinder
Cuff
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...