5-6 VS 9/VS 8 Vital Signs Monitor Service Manual
Fault Symptom
Possible Cause
Troubleshooting
The IP address is
incorrect.
Check whether an IP address conflict exists in
the network. If yes, reset the IP address.
Connector board
damaged
Replace the connector board
The main board is
faulty
Replace the main board
The network
connection is
disconnected
frequently.
The network cable
fails to be
connected
properly.
Check LAN cable connection. LAN cable should
not be longer than 50m.
The monitor is
connected to a
network but
cannot view
other patients
in the View
Others mode
The network cable
fails to be
connected
properly.
Check LAN cable connection. LAN cable should
not be longer than 50m.
There are too many
monitors to be
viewed.
One monitor could only be observed by four
monitors simultaneously, and the observing
requests not within the range would not be
handled.
The IP address is
incorrect.
Check whether an IP address conflict exists in
the network. If yes, reset the IP address.
5.6.11
Wi-Fi Failures
Fault Symptom
Possible Cause
Troubleshooting
Frequent dropouts
or network
disconnects
Wi-Fi signals in the region are
instable.
Check the Wi-Fi connection of
the hospital.
The Wi-Fi- antenna falls off or is
not securely connected to the
module.
Disassemble the monitor and
fix the Wi-Fi antenna properly.
Antenna damaged
Replace the antenna
The Wi-Fi module is faulty
Replace the Wi-Fi module
The monitor fails to
connect Wi-Fi
The IP address is incorrect.
Check whether an IP address
conflict exists in the network.
If yes, reset the IP address.
Wi-Fi signals in the region are
instable.
Check the Wi-Fi connection of
the hospital
The Wi-Fi antenna falls off or is
disconnected from the Wi-Fi
module
Secure the Wi-Fi antenna
Antenna damaged
Replace the antenna
The Wi-Fi module is faulty
Replace the Wi-Fi module
The main board is faulty
Replace the main board
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...