4-10 VS 9/VS 8 Vital Signs Monitor Service Manual
4.5.4
CO
2
Test
4.5.4.1
CO
2
Leakage Test
It tests whether the CO
2
module leaks.
Test Preparations
None.
Procedure
Follow this procedure to perform the test:
1. Start up the monitor. After the CO
2
is warmed up, completely block the gas inlet of
the module or watertrap (you may use a pneumatic plug or your finger to manually
occlude the port).
2. Sidestream CO
2
module: After 3 seconds, the alarm message
CO2 Airway Occluded
is
displayed on the screen. Block the gas inlet for another 60 seconds. Select
Main
Menu
->
Maintenance
-> enter the password ->
Module
->
CO2
->
Calibrate
to check
that the current CO
2
flow rate is less than 10 ml/min. If the alarm message remains
on the screen, it indicates that the module does not leak.
4.5.4.2
CO
2
Accuracy Test
It tests the CO
2
module accuracy.
Test Preparations
Test tools:
Steel gas cylinder with 3%-7%±0.3% CO
2
and balance gas N
2
(P/N:
0075-00-0033-01)
T-shape connector
Air tubing
Procedure
Follow the procedure below:
1. Turn on the monitor. Wait until the CO
2
module warm-up finishes. Check the airway
for leakage and perform a leakage test as well to make sure that the airway has no
leakage.
2. Select
Main Menu
->
Maintenance
-> enter the password ->
Module
->
CO2
->
Calibrate
.
3. Connect the test system as follows:
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...