8-8 VS 9/VS 8 Vital Signs Monitor Service Manual
8.2.3.2
Entering the Upgrade Mode
Method 1: Power off the monitor, and connect the monitor to a USB keyboard.
Press the power key to power on the monitor. During the power-on, keep
pressing F4+F5 at the same time or press * to enter the upgrade mode to
perform upgrade.
Method 2: Power off the monitor. Press the power key to power on the monitor.
During the power-on, keep sliding two or more fingers on the touchscreen to
enable the monitor to enter the upgrade mode.
8.2.3.3
Selecting the File for Upgrade
Only one upgrade package file is available at present. It is selected by default. If multiple
upgrade packages exist, they are displayed in the left and right columns, and a maximum
of 16 upgrade packages can be displayed. This upper limit cannot be exceeded. You can
use the up, down, left and right keys to switch and select the required upgrade package.
Tap
on the touchscreen or press the
arrow key on the
keyboard to switch the upgrade program downward.