4-18 VS 9/VS 8 Vital Signs Monitor Service Manual
Connection Diagram
Connect ESA612
to the ground
power socket.
Connect to protective ground terminal or any
exposed conductive surface on the chassis.
Connect the AC power cable of the device to be
tested to the equipment output of the analyzer.
Figure 9 Connection Diagram
1. Insert the power plug of the analyzer into the socket and turn on the power button
to enter the startup screen.
2. Press the
SETUP
key and press the function key
More
to display other menu options.
Press the function key
Instrument
to display the instrument setting options. Press
the function key
Standard
and select the detection standard (62353).
Figure 10 Equipment Leakage Current Test
Procedure
Equipment leakage current test procedure: