VS 9/VS 8 Vital Signs Monitor Service Manual
6-39
6.4.5.3
(F-3) Removing the NIBP Air Tube
Same as F-2 Removing the NIBP Pump
6.4.6
(G) Sheet Metal Main Support Assembly
Unplug the AC cable socket 1, and remove the sheet metal from bottom to top to
disassemble the assembly into the plastic main support assembly and the sheet metal
main support assembly.